Skip to main content
BrainCited

Bidirectional (positive/negative) interference of spironolactone, canrenone, and potassium canrenoate on serum digoxin measurement: elimination of interference by measuring free digoxin or using a chemiluminescent assay for digoxin.

Amitava Dasgupta, Helene Saffer, Alice Wells, Pradip Datta
Other Journal of clinical laboratory analysis 2002 13 citations
PubMed DOI
<\/script>\n
`; }, get iframeSnippet() { const domain = 'braincited.com'; const params = 'pmid\u003D12112389'; return ``; }, get activeSnippet() { return this.method === 'script' ? this.scriptSnippet : this.iframeSnippet; }, copySnippet() { navigator.clipboard.writeText(this.activeSnippet).then(() => { this.copied = true; setTimeout(() => { this.copied = false; }, 2000); }); } }" @keydown.escape.window="open = false" @click.outside="open = false">

Embed This Widget

Style



      
      
    

Widget powered by . Free, no account required.

Study Design

Type d'étude
In Vitro
Population
Lab assay evaluation
Intervention
Bidirectional (positive/negative) interference of spironolactone, canrenone, and potassium canrenoate on serum digoxin measurement: elimination of interference by measuring free digoxin or using a che Spironolactone, canrenone
Comparateur
None
Critère de jugement principal
Interference with digoxin immunoassay
Direction de l'effet
Negative
Risque de biais
Unclear

Abstract

Spironolactone and potassium canrenoate (aldosterone antagonist diuretics) are often used with digoxin in clinical practice. Spironolactone, potassium canrenoate, and their common metabolite canrenone cross-react with the fluorescence polarization immunoassay (FPIA) for digoxin, and can falsely elevate serum digoxin concentrations. Serum digoxin concentrations were falsely lowered when the microparticle enzyme immunoassay (MEIA) was used. Aliquots of drug-free serum were supplemented with therapeutic and above-therapeutic concentrations of spironolactone, canrenone, and potassium canrenoate, and apparent digoxin activities were measured. We observed digoxin-like activities in the FPIA, but observed no activity with the MEIA or the chemiluminescent assay (CLIA). However, when serum digoxin pools prepared from patients receiving digoxin were supplemented with these compounds, we observed suppression of total digoxin levels with the MEIA. In contrast, no interference was observed in the presence of these compounds when CLIA was used for digoxin measurement. These compounds are strongly protein-bound, and no apparent digoxin activity was observed in the protein-free ultrafiltrate when drug-free sera were spiked with high levels of these compounds. Taking advantage of strong protein binding of these compounds and weak protein binding of digoxin (25%), interference of spironolactone, canrenone, and potassium canrenoate in FPIA and MEIA digoxin assays can be mostly eliminated by monitoring free digoxin concentration. Another approach to avoid this interference is to use the CLIA digoxin assay.

En bref

Interference of spironolactone, canrenone, and potassium canrenoate in FPIA and MEIA digoxin assays can be mostly eliminated by monitoring free digoxin concentration.

Used In Evidence Reviews

Similar Papers